Specifications / Standard Specifications

Standard Specifications

Standard Specifications
Construction Mounting Flush panel mounting (on a vertical plane) Mounting may be inclined downward up to 30 degrees from a horizontal plane.
Allowable panel thickness 2 to 26 mm
Front panel Water and dust-proof (based on IEC529-IP65 and NEMA No.250 TYPE4)
Input Number of inputs DX1000 2, 4, 6, 12 channels
DX2000 4, 8, 10, 20, 30, 40, 48 channels
Measurement interval DX1002, DX1004, DX2004, DX2008 125 ms, 250 ms, 25 ms (fast sampling mode)
DX1006, DX1012, DX2010, DX2020, DX2030, DX2040, DX2048 1 s (Not available when A/D integration time is set to 100 ms), 2 s, 5 s, 125 ms (fast sampling mode*)
* A/D integration time is fixed to 1.67 ms in case of fast sampling mode.
Inputs DCV 20, 60, 200 mV, 2, 6, 20, 50 V, 1-5 V
TC R, S, B, K, E, J, T, N, W, L, U, WRe
RTD Pt100, JPt100
DI Contact input, TTL level
DCA With external shunt resistor attached
Input Range Measurement accuracy (when the integration time is 16.7 ms or more) Display resolution
DCV 1-5 V ±(0.05% of rdg+3 digits) 1 mV
Thermocouple*
K
* Does not include the accuracy of reference junction compensation
±(0.15% of rdg+0.7℃) 0.1°C
Resistance thermometer detector Pt100 ±(0.15% of rdg+0.3℃) 0.1°C
Display Display unit DX1000 5.5-inch TFT color LCD (320 x 240 pixels)
DX2000 10.4-inch TFT color LCD (640 x 480 pixels)
Display group Number of display DX1000: 10 groups, DX2000: 36 groups
Number of assignable channels for one group DX1000: 6 channels, DX2000: 10 channels
Display color Trend/Bargraph Selectable from 24 colors
Background White or black selectable
Trend display Trend display type Vertical, horizontal, landscape, horizontal split or circular* selectable
* Circular display is only for DX2000.
Bargraph display Direction Vertical or horizontal selectable
Digital display Display renewal rate 1 s
Overview display Number of indication channels Measuring values and alarm status of all channels
Information display Alarm summary, message summary, memory information, report information, relay status, Modbus status
Tag display Number of displayable characters Tag number: Max. 16 characters
Tag comment: Max. 32 characters
Displayable characters Tag number: Alphanumerics
Tag comments: Alphanumerics, Japanese, and Chinese
Message display Number of characters Max. 32 characters
Number of messages 100 messages (including 10 free messages)
Historical display function Display the retrieved data (display data or event data) from internal or external memory.
Custom display function
User can change display object (trend, numeric, and bar graphs, etc.) sizes and attributes, and add objects freely to create screens.
Number of screens 28 (3 from internal memory, 25 from external media (CF))
Max. number of
placeable
display objects
132 (normal: 80, scale: 4, trend: 4, list: 4, graphic: 40)

Data Saving Function External storage medium Medium CompactFlash memory card (CF card)
Internal memory Medium Flash memory
Capacity 400 MB
Max. number of files can be saved 400 files (total number of display data file and event data file)
Alarm Function Number of alarm levels Up to 4 levels for each channel
Alarm types High and low limits, differential high and low limits, high and low rate-of-change limits and delay high and low
Alarm annunciator function Alarm display based on alarm sequence, and relay output operation. Supported alarm sequences: 3 (ISA-A-4, ISA-A, ISA-M)
Event Action Function General Particular action can be executed by particular event.
Number of event action 40 actions can be set
Security Functions General *
* Please refer the Advanced security function option (/AS1) for the models with the /AS1 option.
Login function or key lock function can be set for each key operation or communication operation.
Key lock function On/off and password can be set for each operation key and FUNC operation.
Login function User name and password to login can be set.
Clock Clock With calendar function (year of grace)
Clock accuracy ±10 ppm, excluding a delay (of 1 second, maximum) caused each time the power is turned on.
DST function (summer/winter time)
The time at which the daylight savings time adjustment is automatically calculated and configured.
Communication Function Connection Ethernet (10BASE-T)
Protocols TCP, UDP, IP, ICMP, ARP, EtheNet/IP, DHCP, HTTP, FTP, SMTP, SNTP, Modbus, DX private
Communication function E-mail function, FTP client function, Transferred data file, FTP server function, Web server function, SNTP client function, SNTP server function, DHCP client function, Modbus client function, Modbus server function
EtherNet/IP server Connects to EtherNet/IP networks as an adapter (server).
Batch Function Overview Data display and data management with batch name, text field function and batch comment function are available.
Power Supply Rated power supply 100 to 240 VAC (automatic switching)
Allowable power supply voltage range 90 to 132 or 180 to 264 VAC
Rated power supply frequency 50/60 Hz(automatic switching)
Power consumption             DX1000 60 VA (max., for 240 VAC power supply)
DX2000 100 VA (max., for 240 VAC power supply)
Normal Operating Conditions Power voltage 90 to 132 or 180 to 250 VAC
Power supply frequency 50 Hz ±2%, 60 Hz ±2%
Ambient temperature 0 to 50 °C
Ambient humidity 20% to 80% RH (at 5 to 40 °C)
| Print Page |